Alcotest 7410 Med
Draeger

Alcotest 7410 Med


The Alcotest 7410 med is a special version of
the proven Dräger Alcotest series. It has been designed for diagnostic applications, where the CE marking for medical products is necessary.
In addition to the electrochemical DrägerSensors quality and proven reliability, the Dräger Alcotest 7410 med offers a variety of features that ease daily practices within the medical field.

Application areas

The Alcotest 7410 med has been adapted to determine breath alcohol concentrations     especially in medical applications. It may be used, for example, to diagnose vascular injuries (about others in the urology during transurethral resections of the prostate) after previous application of an ethanol containing rinsing solution. It is also used for testing patients for previous alcohol consumption at the emergency reception area or before methadone substitution.

Features

The speed and long-term stability of the DrägerSensor leads to a short waiting time
for the first measurement after the instrument
has been turned on, a fast determination of the measured value and a small standard deviation. The illuminated instrument display with
information such as test number, date, time, measurement result and error messages as
well as the coloured LEDs and acoustical
alarms guide the operator easily through the
use of the instrument.

Sampling


During the automatic measurement, which is activated by the breath sample, the breath flow is monitored and the minimum volume required is adjusted to ensure deep lung air is taken. The manually activated measurement makes measurements possible even with people who are not able to deliver intentionally a breath sample.






Alcotest 7410 Med
Alcotest 7410 Med



Data storage

The integrated data memory can store approximately 8000 data sets for person
related measurement results. Via the interface, information such as the patient and the place of measurement can be downloaded into the instrument before the measurement and the results can be transferred into a PC after the measurement.

CE marking as Medical Device The Alcotest 7410 med has a CE marking according to the EU Directive 93/42/EEC for Medical Devices. According to this legislation, this CE marking is necessary for all applications involving “diagnosis, prevention, monitoring, treatment or alleviation of disease”. The Alcotest 7410 med fulfils these requirements for applications such as during operations, at emergency reception areas or during methadone substitution.


PDF

Alcotest 7410 Med Brochure

Alcotest 7410 Med Manual


Alcotest 7410 Med Specification


Measuring Principle Electrochemical sensor
Measurement range 0 to 1.5 mg/L
Display range 0 to 9.99 mg/L
Sample conditions
(automatic measurement)
Blowing time: 2 to 12 s
(depends on the blowing intensity)
Breath flow: > 6 L/min
Breath volume: > 0.3 L
Measurement accuracy
(reproducibility with an
ethanol standard
0 to 0.5 mg/L: ± 0.03 mg/L
> 0.5 mg/L: ± 5 % of measured value
Sensitivity drift typically 0.6 % of measured
value per month
Cross sensitivity to
anesthetic gases
none
Calibration interval 6 months
Function test interval weekly (recommended)
Dimensions Mw x oh x d 70 mm x 230 mm x 34 mm
Weight apron. 500 g
Electrical supply 3 x 1.5 V (type LR 14)
Operation time at
room temperature
apron. 400 measurements or
8 oh continuous operation per battery set
Interface RS 232 interface for data transmission to a PC
Classification as per
Directive 93/42/EEC,
Annex IX
Measuring instrument: Class I with
measuring function Mouthpieces: Class I
UMDNS Code 17-475



ISO 9001
More Information
UK UK Office
Tel: +44 (0)1245 600 560
Fax: +44 (0)1245 808 399
Email: info@keison.co.uk
USA USA Office
Tel:
+1 541 956 9929
Fax: +1 541 610 1646
Email: info@keison.com
Germany German Office
Tel:
+49 761 4787 184
Fax: +49 761 4787 100
Email: info@keison.de
ISO 140001
More Information


© Copyright 2008 Keison International Ltd - All Rights Reserved