Alcotest 7410 Plus
|
The Alcotest 7410 product family is the most successful hand-held breath
alcohol measuring instrument. Used on all five continents, for control
of alcohol at workplaces in industry, for monitoring or medical diagnosis
of patients and for traffic control by police authorities. More than 50,000
Alcotest 7410 are in use worldwide currently.
Features and Benefits of the Alcotest 7410
Plus:
quick and long-term
stable DrägerSensor: short waiting time until first test, quick determination
of the measured value, quick retest rate and small standard deviation
low blowing resistance
real-time clock
with date and time
measurements are
also possible without cooperation of the person being tested by passive
sampling: triggered by pressing a button and afterwards qualitative detection
of alcohol
data memory for
more than 9,700
measurement results with date and
time information: the stored data may be
recalled menu-driven by using a button on
the instrument
statistical options
protocol print-out
by the Alcotest 7410 Printer
three different
power supply units: with
alkaline batteries (up to 600 measurements),
with rechargeable accumulator (up to 300 measurements) or motor vehicle
connection 12 V
charging station
with indication of
charging condition and overcharging protection
for accumulator power supply unit.
The Alcotest 7410Plus RS offers you:
the features and
benefits of the Alcotest 7410 Plus
|

Alcotest 7410 Plus
and additionally:
a data interface
(RS 232) for
connection to a PC
an input of person-specific
data for a test and read-out of the up to 9,700 stored data sets with
the Dräger PC Software “View 2000“
evaluation, print-out
and management of stored data.
The exceptional long-term stability of the DrägerSensor and the specific
sampling system allows long calibration intervals for the Alcotest 7410
Plus and Alcotest 7410 Plus RS within the quality assurance. Several calibration
systems are available for calibration. In addition you can let the calibration
and care of your Alcotest 7410 be performed by our DrägerService.
Alcotest 7410 Plus Brochure
Alcotest 7410 Plus Manual
|
|
Alcotest
7410 Plus Specification
| Measuring Principle |
Electrochemical Sensor |
| Measuring Unit |
Mg/L % (further measurement units available) |
| Measurement Range |
0.00 to 1.50 mg/L; 0.00 to 3.00 %, 0.000 to 0.300 % |
| Display Range |
0.00 to 9.99 mg/L, 0.00 to 9.99 %, 0.000 to 0.999 % |
| Ambient Conditions |
-5 to 45°C /23 to 113°F
(and at -20°C / -4°F but not for more than 30 minutes and
only after being kept above 0°C /32°F until use)
For storage -40 to 65°C /-40 to 149°C
|
| Sampling |
Blowing: 4-12 s (depends on the intensity
Breath flow: > 6L /min
Breath Volume: > 1.2L (Adjustable |
Measurement Accuracy
(max.error of repeatability within an ethanol standard) |
0-0.5 mg/L : ± 0.03 mg/L
0-1% ± 0.05%
0.5 mg/L: ± 5% of measured value |
| Sensitivity drift |
Typically 0.6% of measured value/month |
| Intervals before readiness of operation |
1st measurement: approx 15 s after determining a previous test result
in range:
Up to 0.19 mg/L: 10 s >0.95 mg/L: 90 s |
| Calibration interval |
recommended 6 months (depends on required accuracy) |
| Dimensions |
70mm x 230mm x 34mm (w x h x d) |
| Weight |
Approx 500g |
|
 |
|
|
|
| Site Map | Enquiry Form | Translate |