| Specification |
|
| Gases Measured: |
O2 |
| PERFORMANCE |
|
| Range: |
0-100% O2 |
| Output: |
Two 4-20mA, isolated & one 0-1V dc, non-isolated. Ranges selectable from 0 - 2.5, 5, 10, 25 and 100% O2 |
Alarm: Concentration
|
Two volt free changeover relays |
| Alarm: Sample flow fail |
One volt free changeover relay |
| Response time (T90) |
< 12 seconds at recommended sample and bypass flow rates* |
| Noise: |
< 0.1% O2 at optimum bypass flow rate with Nitrogen |
| PHYSICAL |
|
| Dimensions: (W x D x H) |
448mm x 229mm x 235mm (17.6" x 9" x 9.25") |
| Inlet Pressure: |
2 to 140kPa (0.3 to 20 psig) |
| Sample Temperature: |
Suitable for use with samples saturated at up to +45°C (+113°F) |
| Sample flow rate: |
200ml/min recommended |
| Bypass flow rate |
400 to 1100 ml/min (air) |
| Filter Retention: |
90% of 0.6µm particles |
| Inlet/Outlet Connections: |
1/4" NPT (Int) Connectors |
| Operating Temperature |
-10°C to +50°C (+14°F to 122°F) |
| Storage Temperature |
-10°C to +50°C (+14°F to 122°F) |
| Atmospheric Pressure: |
79 to 124kPaa/11 to 18psia |
* Includes Xendos 1800 response time
Below 0°C the water in the bubbler should be replaced by ethylene glycol |