| TECHNICAL
DETAILS |
| Input pressure |
Up to 150 psig, 10bar 100kpa |
| Over pressure |
Over range protection: 'OUT' also shows on display |
| Pressure Media |
Dry non corrosive air or gas. The units are not suitable
for continuous liquid exposure |
| Display |
12.7mm/0.5" LCD |
| Power |
2* AA (or equivalent) batteries |
| Consumption |
Batteries last approx 200 hours |
| Low battery voltage |
Display shows symbol at approximately 6.5V |
| Auto Switch-off |
Approximately 12 minutes |
| CONSTRUCTION
|
| Casing |
Moulded plastic, black finish |
| Dimensions |
150*80*30mm |
| Weight |
180g (inc. battery) |
| Max/Min: |
Enables max/min pressures to be determined, as well as current
operating pressure |
| Hold |
Freezes the reading |
| Unit |
Choice of 10 units of pressure |
| Zero |
To Zero the instrument in the set reading orientation |
| Range Lock |
To set the instrument to the lower resolution only |
| Smooth |
Averages measurements for a more stable reading |
| ACCURACY
|
| Calibration |
+/- 0.1% FS
|
| Hysteresis / Linearity |
+/- 0.15% FS +/- 1 digital total |
Temperature Coefficients
- Zero
- Span |
+/0.0025% FS/C Max
+/0.0025% FS/C Max |
| Stability over 6 months |
+/- 0.2% Maximum |
| Operating Temperature |
-10 to 50°C |
| Storage Temperature |
-20°C to +60°C |
| Note |
Figures given are typical and not absolute for all units
manufactured |
| CERTIFICATION
|
|
All units are individually tested and a certificate is supplied
with every unit
Type 300D Brochure
|