Furse cabinet dimensions (mm)* (h x w x case depth x overalldepth/including depth of door) | |
Work tube (internal diameter mm) | 79 |
Maximum temperature degC (degF) | 1600 (2912) |
Maximum recommended heat up rate degC/minute (degF/minute) | 8 (15) |
Temperature control | Digital PID with multi offset-parameters |
Temperature sensor | Pt & Pt/13% Rh thermocouple |
Overtemperature protection | Digital with high alarm relay |
Power supply | 380/415V, 50/60hz, two phase, 25A/phase or 220 x 240V, 50/60Hz, single phase, 50A Other voltages available on request |
Maximum power consumtion (kW) | 7 |
Gases: Two options are available for use with the following gases. Gas option has to be specified with order. | |
Reducing gas Oxidising gas Purge gas | 60% (V/V) CO + 40% (V/V) CO² Air N² |
or reducing gas Oxidising gas Purge gas | 50% (V/V) H² + 50% (V/V) CO² CO² CO² (For use with H²) |