1200°C Tube Furnaces GHA, GVA, GHC, GVC
|
|

The carbolite G range of tube furnaces replaces the GTF/GZF furnace range.
With a maximum operating temperature of 1200°C, these furnaces can accept work tubes with a diameter from 19.5 — 150nmi. The range includes seven heated lengths from 300mm to 1200mni in both horizontal and vertical versions, with either single-zone or three-zone control. The multi-zone models provide particularly good temperature uniformity and can be used to create a temperature gradient.
Heating is provided by semi-embedded, free-radiating wire elements in vacuum formed insulation modules, which provide good support and increase the working life of the element. An integral work tube is not fitted and a separate work tube must be used, but removable tube adaptors at each end of the furnace allow rapid changes for different size work tubes.
|
 
The horizontal versions are mounted on an attractive base unit, whilst the versatile models are supplied with a stand kit and a separate control box with a length of conduit flex which permits horizontal, vertical and wall mounted use.

Carbolite 1200C Tube Furnaces Specification
|
The model number listed in the table below must be prefixed by one of the following, eg GVA 12/450.
GHA = 1 zone horizontal
GVA = 1 zone versatile (with multiple use stand)
GHC = 3 zone horizontal
GVC = 3 zone versatile (with multiple use stand)
| Model |
12/300 |
12/450 |
12/600 |
12/750 |
12/900 |
12/1050 |
12/1200 |
| Max Temp (°C) |
1200 |
| Optimum Temp Control (°C) |
800 - 1200 |
| Heated Length (mm) |
300 |
450 |
600 |
750 |
900 |
1050 |
1200 |
| Work tube diameter (mm) |
19.5 - 150mm |
Work tube length
Type C
Type D |
480
980 |
630
1130 |
780
1280 |
930
1430 |
1080
1580 |
1230
1730 |
1380
1880 |
External dimensions (Horizontal model - mm) H
W
D |
670
526
468 |
670
676
468 |
670
826
468 |
670
976
486 |
670
1126
468 |
670
1276
468 |
670
1426
468 |
Height on stand to top of furnace. Maximum
Adjustment |
1342 |
1418 |
1418 |
1660 |
1660 |
1660 |
2018 |
| Stepped 37.5mm from maximum to minimum |
Clarence under furnace
Maximum
Minimum |
777
252 |
702
177 |
550
177 |
777
177 |
702
102 |
627
101 |
551
101 |
Remote control box dimensions
(h x w x d mm) fitted 2 m conduit cable |
222 x 570 x 375 |
Electrical supply:
220 - 240V single phase
380 - 220V - 415/240V 3 phase + neutral
208 - 240V 3 phase delta |
yes
no
no |
yes
yes
yes |
yes
yes
yes |
yes
yes
yes |
yes
yes
yes |
yes
yes
yes |
yes
yes
yes |
| Carbolite |
| Ovens and Incubators | Furnaces | High Temperature Furnaces | Tube Furnaces |